ACT Kaolin |
ACT Kaolin |
Used to monitor high dose heparin anticoagulation therapy
40μL
Arterial or venous whole blood
03P87-25
Reportable Range | Reference Range (Arterial) | Reference Range (Venous) | |
ACT Kaolin | 50-1000 seconds | 74-137 seconds (Prewrm) | 74-137 seconds (Prewrm) |
ACT Celite |
ACT Celite® |
Used to monitor moderate and high level heparin therapy
40μL
Arterial or venous whole blood
03P86-25
Reportable Range | Reference Range (Arterial) | Reference Range (Venous) | |
ACT Celite® | 50-1000 seconds | 74-125 seconds (Prewrm) | 74-125 seconds (Prewrm) |
With i-STAT ACT, healthcare professionals can measure the time required for complete activation of the coagulation cascade. i-STAT ACT is similar to traditional ACT tests except that the endpoint is indicated by a chemical measurement of the presence of thrombin instead of a mechanical measurement of a physical clot. i-STAT ACT provides clinicians a with-patient ACT test that is more reliable and reproducible, enabling more accurate heparin dose adjustments.
In a study that compared the precision and reliability of i-STAT Kaolin ACT with Hemochron Signature Elite and Medtronic ACT Plus, i-STAT ACT demonstrated better performance and was less susceptible to external factors such as temperature, dilution, hematocrit, fibrinogen levels and user technique.
For an overview of i-STAT System test cartridges, browse the Test Cartridge Menu.
Providing lab-quality results fast when every minute counts.
Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott. The website that you have requested also may not be optimised for your screen size.
Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions.