Crea Cartridge

About CREA

The i-STAT Creatinine test is used for the quantitative determination of creatinine (crea) in whole blood on the i-STAT handheld. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

This basic metabolic panel for use in the in vitro quantification of creatinine in whole blood samples allows healthcare professionals to obtain results in approximately two minutes to detect elevated levels of creatinine mainly associated with abnormal renal function occurring whenever there is a significant reduction in glomerular filtration rate or when urine elimination is obstructed.

Imaging agents (such as iodine and barium) used in radiological procedures can pose serious health risks to patients with certain conditions. To minimize risk, creatinine measurement of renal function is recommended prior to administering imaging agents. Because it may take several hours for the clinical lab to process creatinine tests, i-STAT Crea tests can be used to minimize disruptions and improve efficiencies.

In an article published in British Journal of General Practice, i-STAT Crea test values showed excellent correlation (R² = 0.99) compared with core lab values, with the authors concluding that the i-STAT System could help improve efficiency, providing accurate, up-to-date creatinine/eGFR values in patients presenting for scans requiring contrast media.1

For an overview of i-STAT System test cartridges, browse the Test Cartridge Menu.

Reference: 1. Creatinine Point-of-Care Testing for Detection and Monitoring of Chronic Kidney Disease: Primary Care Diagnostic Technology Update, Oghenekome Gbinigie, Christopher P Price, Carl Heneghan, Ann Van den Bruel, and Annette Plüddeman, British Journal of General Practice, November 2015.

130_x_160_Time_to_Result_Crea.png

Results in ≈2 Minutes

Providing lab-quality results fast when every
minute counts.

e-Abbott

Interested in Crea cartridges? Order Crea cartridges through our e-Abbott portal.

Order Crea

Cartridge Details
 

chemistries_electrolytes.pngChemistries/Electrolytes
 
Crea
Creatinine (Crea)

 

Crea

Description:

Rapid results for assessing renal function

Sample Size:

65μL

Sample type:

 

WAIVED:

Venous whole blood samples collected in lithium heparin evacuated tubes (green top tubes) when run on the i-STAT 1 System

Moderate
complexity:

Arterial, capillary, or other venous whole blood sample

Part Number:

03P84-25

Expected Values
 

chemistries_electrolytes.pngChemistries/Electrolytes
 
Reportable Range Reference Range (Arterial) Reference Range (Venous)
Creatinine (Crea) 0.2-20.0 mg/dL 0.6-1.3 mg/dL 0.6-1.3 mg/dL

 

You are about to exit the Abbott family of websites for a 3rd party website

Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott. The website that you have requested also may not be optimised for your screen size.

Do you wish to continue and exit this website?

Yes No

You are about to enter an Abbott country or region specific website.

Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions.

Do you wish to continue and enter this website?

Yes No

Not optimized for your device

The page you are trying to access is best viewed on a desktop or laptop computer.

I want to view this page anyway
Contact us today